BECKMAN COULTER Centrifugation & Ultracentrifugation
The guarantee of satisfaction throughout the life of your investment.
Analis offers the following services linked to centrifugation:
- Prevention Program - FRIP: Field Rotor Inspection Program
- Training for best user practices
- Advice on the equipment choice and for parameter optimization
- Advices for balancing your rotor
Don't hesitate to contact our Technical Service : TEL + 32 81 25 50 50 - +32 9 243 77 10
Download brochure of our technical service
Analis offers the following services linked to centrifugation:
- Prevention Program - FRIP: Field Rotor Inspection Program
- Training for best user practices
- Advice on the equipment choice and for parameter optimization
- Advices for balancing your rotor
Don't hesitate to contact our Technical Service : TEL + 32 81 25 50 50 - +32 9 243 77 10
Download brochure of our technical service
*Prevention Program - FRIP: Field Rotor Inspection Program
Analis offers to the users of Beckman Coulter centrifuges and ultracentrifuges a rotor inspection program to ensure longevity of equipment.
These inspections prevent accidents and prevent premature damage on rotors due to:
A report on the condition of the rotors and the possible limitations of use is sent to you at the end of the inspection.
If you wish to benefit from one of these services or to receive more information, contact us.
*Trainings
a) Analis Academy - Education session centrifugation
b) Training for best user practices can be organized in your or our facilities.
Objectives of the outdoor trainings:
Analis offers to the users of Beckman Coulter centrifuges and ultracentrifuges a rotor inspection program to ensure longevity of equipment.
These inspections prevent accidents and prevent premature damage on rotors due to:
- Stress corrosion
- The fatigue of the material
- Wear or damage to the coating
A report on the condition of the rotors and the possible limitations of use is sent to you at the end of the inspection.
If you wish to benefit from one of these services or to receive more information, contact us.
*Trainings
a) Analis Academy - Education session centrifugation
b) Training for best user practices can be organized in your or our facilities.
Objectives of the outdoor trainings:
- Safe centrifuge use
- Maximum lifetime warranty of your investment
- Avoiding unbudgeted expenses
- Proper use and handling of rotors
- Good use of tools
- Care and maintenance of rotors
- Optimal choice of tubes
- After sales support
- Out of warranty for rotors and decommissioning
*Advice on the choice of rotors / tubes / accessories and optimization of centrifugation parameters
- Our centrifuge product ranges
- Intellifuge Beckman Coulter rotor calculator
*Advices to balance your rotor
In order for a rotor to run smoothly and safely at its operating speed, the load it carries must be balanced. A rotor can be properly balanced by following some simple rules:
You may notice that the centrifuge vibrates when the rotor is accelerating or decelerating at low speeds. This is normal, and occurs as the rotor passes through a so called critical speed range where any small vibrations are temporarily amplified. Your separations will not be disturbed during deceleration, because the centrifugal force is still high enough to stabilize them. However, you should not select an operating speed within the range where these exaggerated vibrations occur. Your instruction manual will tell you what speeds to avoid.
Correct Symmetry When Balancing a Partial Load – A Top View of a Horizontal Rotor
Example of a balanced load:

Example of an unbalanced load:

Even if all the tubes are filled equally, this rotor is improperly loaded. None of the bucket loads are balanced with respect to their pivotal axes. At operating speed, buckets A and C will not reach the horizontal position. Buckets B and D will pivot past the horizontal. Also note that the tube arrangement in the opposing buckets B and D is not symmetrical across the center of rotation.
Don't hesitate to contact our Technical Service : TEL + 32 81 25 50 50 - +32 9 243 77 10
Téléchargez notre brochure du service technique - Download onze brochure van de technische dienst
- Our centrifuge product ranges
- Intellifuge Beckman Coulter rotor calculator
*Advices to balance your rotor
In order for a rotor to run smoothly and safely at its operating speed, the load it carries must be balanced. A rotor can be properly balanced by following some simple rules:
- A rotor must never be run with buckets missing, although opposing buckets may be left empty.
- All opposing loads must balance within a certain weight as specified by the centrifuge manufacturer’s instruction manual.
- If opposing buckets are run with a partial load of tubes in their adapters, these tubes must be arranged symmetrically, both with respect to the pivotal axis of each bucket and across the center of rotation. With some partial loads, it may be difficult or impossible to achieve the correct symmetry in both sets of opposing buckets. The simplest solution is to fill one or more tubes of the same size with water, or a denser liquid if necessary, and use them to balance the load symmetrically.
You may notice that the centrifuge vibrates when the rotor is accelerating or decelerating at low speeds. This is normal, and occurs as the rotor passes through a so called critical speed range where any small vibrations are temporarily amplified. Your separations will not be disturbed during deceleration, because the centrifugal force is still high enough to stabilize them. However, you should not select an operating speed within the range where these exaggerated vibrations occur. Your instruction manual will tell you what speeds to avoid.
Correct Symmetry When Balancing a Partial Load – A Top View of a Horizontal Rotor


Example of an unbalanced load:
Even if all the tubes are filled equally, this rotor is improperly loaded. None of the bucket loads are balanced with respect to their pivotal axes. At operating speed, buckets A and C will not reach the horizontal position. Buckets B and D will pivot past the horizontal. Also note that the tube arrangement in the opposing buckets B and D is not symmetrical across the center of rotation.
Don't hesitate to contact our Technical Service : TEL + 32 81 25 50 50 - +32 9 243 77 10